Hemochron ACT (Activated Clotting Time) analysers

 

POCT Test Name

Activated Clotting Time 

 

Instruments

Hemochron 401

Hemochron Response

Hemochron Elite

Hemochron 401

Hemochron Signature Elite

 

Location within CDHB

Radiology DSA, Christchurch Hospital - 401 model.

Cardiac Cath Lab, Christchurch Hospital- Response model.

Operating Room -Theatre 9, Christchurch Hospital- Signature Elite.

 

Tests measured

Heparin monitoring with or without Aprotinin.

 

Sample and volume requirements

Fresh whole blood- 2mL. Elite: 200µL to fill cuvette well.

 

Consumables

401/Response

FTCA510/HRFTCA510- regular testing

FTK-ACT/HRFTK-ACT- Aprotonin use only

 

Signature Elite

State-of-the-art features like barcode scanning, internal Electronic Quality Control (EQC), QC Lockout and alphanumeric data entry for Patient Identification (PID) and Operator Identification (OID).

 

Controls

Electronic QC

Liquid quality control- normal and abnormal

 

Reference Ranges

 

Normal range- healthy donors

RFTCA 510 tubes: 105 – 167 seconds

FTK-ACT tubes: 91 – 151 seconds

 

Normal range- cardiopulmonary bypass patients

HRFTCA 510 tubes: 86 - 147 seconds

FTK-ACT tubes: 89 - 153 seconds

 

ECMO therapeutic ranges

220 – 260 seconds, varies according to patient diagnosis

 

Paediatric dialysis, therapeutic ranges

180 – 220 seconds

 

(Reference Ranges are taken from the Hemochron package insert)

 

Specimen Rejection 

Tissue thromboplastin

Indwelling intravenous (IV( solutions

Alcohol cleansing solutions

Visible clots or debris

 

Competencies

Available. Please refer to the POCT Coordinator

 

Return to previous page